DL Series - Dual-Line Analog Input Process Meter
DL-AC-S4NE
The Viatran DL meter boasts specifications and functionality that make it one of the most advanced process meters available.
Viatran's Model 345 High Range pressure sensor is among our most accurate pressure transmitters for industrial test and research applications at higher pressures. Improved accuracies are available, as well as improved temperature performance.
Configure your own product according to your requirements and needs. If you are unsure of the configuration, provide us in the notes the necessary information about the existing sensor or product, even a competitor. We will be happy to help you find an alternative Viatran product. If your request is urgent, let us know. We will do our best to help you resolve this situation to your satisfaction as quickly as possible. If we have all the information, we are able to offer a model with minor differences, which is in stock and you can have another product available the next day.
If you need more information do not hesitate to contact us.
Full Scale Pressure Ranges:
|
7500,10K, 15K, 20K, 25K, 30K, 50K, 60K, 75K, 100K PSIA or PSIG
|
|
Nonlinearity (Best Fit Straight Line):
|
≤±0.1% FSO (≤±0.25% FSO for 100K PSI)
|
|
Repeatability:
|
≤±0.03% FSO
|
|
Hysteresis: |
≤±0.06% FSO | |
Full Scale Output (FSO): |
16 mA at 70°F (21°C) | |
Long Term Stability: | ≤±0.2% FSO per 6 months | |
Compensated Temperature Range: | 70°F to 170°F (21°C to 76°C) | |
Process Media Temperature Range: | -40°F to 250°F (-40°C to 121°) | |
Ambient Operating Temperature Range: | -40°F to 185°F (-40°C to 85°) | |
Storage Temperature Limits: | -40°F to 250°F (-40°C to 121°C) | |
Temperature Effect on Zero: | ≤±0.01% FSO per 1°F - see options | |
Temperature Effect on Span: | ≤±0.01% FSO per 1°F - see options |
Supply Voltage:
|
9.5 to 40 Vdc
|
|
Power Supply Regulation:
|
≤±0.01% FSO per volt change over the supply voltage reange
|
|
Output Signal:
|
4 to 20 mA
|
|
Load Resistance: | 1500 Ohms max at 40 Vdc | |
Zero Adjustment: | ± 10% FSO min | |
Span Adjustment: |
± 10% FSO min |
|
Calibration Signal: | 80% of FSPR, by shorting pins | |
Calibration Signal Accuracy: | ≤±0.1% of the stated value | |
Circuit Protection: | Input polarity may be reversed. Output may be short-circuited indefinitely Over voltage protection to 1000 volts for ≤1 mSec | |
Insulation Resistance: | ≥ 200 MegOhms to case ground | |
Response Time: | ≤ 4.0 mSec | |
Electrical Connection: |
Bendix PT02E-10-6P, |
|
Pin Outs: |
Pin A + Power / Signal Pin B - Power / Signal Pin C Calibrate Pin D Calibrate Pin E No Connection Pin F No Connection |
Wetted Parts
|
7.5K to 15K PSI (500 to 1000 bar)
20K to 100K PSI (1400 to 6900 bar) |
15-5 PH Stainless Steel PH 13-8 MO SST |
|
Housing
|
304 SS with an Aluminum alloy, black zinc-cobalt plated electrical connector
|
||
Weight
|
15 oz (425 g)
|
||
Identification | Laser etched onto body |
Pressure Connection:
|
7.5K to 15K PSI (500 to 1000 bar)
20K to 50K PSI (1400 to 3500 bar) 60K to 100K PSI (4000 to 6900 bar) |
1/4" - 18 NPT female 1/4" F250-C high pres.tube 5/16" F312-C high pres. tube |
|
Proof Pressure:
|
7.5K to 15K PSI (500 to 1000 bar)
20K to 100K PSI (1400 to 6900 bar) |
1.5 times range or 20K PSI (1400 bar) max 1.2 times range or 100K PSI (6900 bar) max |
|
Burst Pressure:
|
7.5K to 20K PSI (500 to 1400 bar)
25K to 100K PSI (1700 to 6900 bar) |
3 times range or 40K PSI (2800 bar) max 1.5 times range or 125K PSI (8600 bar) max |
|
Pressure Cavity Volume: |
7.5K to 15K PSI (500 to 1000 bar) 10K to 100K PSI (1400 to 6900 bar) |
0.32 cubic in (5.2 ml) |
|
Mounting: |
May be supported by process piping |
No Options:
|
No options
|
DL-AC-S4NE