DL Series - Dual-Line Analog Input Process Meter
DL-AC-S4NE
The Viatran DL meter boasts specifications and functionality that make it one of the most advanced process meters available.
The Model 595 Detachable Level Transmitter allows for a quick sensor swap without the removal of the installed cable. No special tools are required. Simply pull up the sensor, twist the sensor coupling sleeve till it is no longer engaged and then simply separate the two halves.
595 is constructed of a PVC housing and ceramic sensor module, with a polyurethane cable as standard. Different cable materials and o rings are available.
Use it to measure Liquid Level and Depth in applications where resistance to caustic materials is required. Semi flush ceramic sensor aids to reduce clogging if media contains solid materials.
Viatran also offers a complete range of accessories for use in level applications such as a surge protector, cable hangers and weights, meters and junction boxes.
Replacement sensor heads and/or cables can be purchased separately
Configure your own product according to your requirements and needs. If you are unsure of the configuration, provide us in the notes the necessary information about the existing sensor or product, even a competitor. We will be happy to help you find an alternative Viatran product. If your request is urgent, let us know. We will do our best to help you resolve this situation to your satisfaction as quickly as possible. If we have all the information, we are able to offer a model with minor differences, which is in stock and you can have another product available the next day.
If you need more information do not hesitate to contact us.
Pressure Ranges | 0 - 16" W.C. up to 0 - 335 Ft W.C. (Or equivalent in Meters or PSI | |
Accuracy [Combined Non-Linearity (BFSL), Hysteresis & Repeatability] | ≤ +/-0.35% FSO (Standard) ≤ +/-0.25% FSO (Optional |
|
Compensated Temperature Range | 32° to 122° F (0 to 50° C) | |
Operating Temperature Range | 14° to 122° F (-10 to 50° C) Note: Medium must not freeze near sensor | |
Storage Temperature Range | 14° to 122° F (-10 to 50° C) | |
Long Term Stability (%FSO) | ≤ +/- 0.1% FSO / year (At reference conditions) | |
Thermal Effect on Zero | ≤ +/- 0.1% FSO per 100° F | |
Thermal Effect on Span | ≤ +/- 0.1% FSO per 100° F |
Output Signal
|
4-20 mA
|
|
Supply Voltage
|
10 - 32 Vdc
|
|
Power Supply Regulation
|
< +/- .005% FSO per volt
|
|
Circuit Protection | Reverse polarity protected, CE marked | |
RFI / EMI | CE EMC compliant as per IEC EN 61326 | |
Voltage Spike Protection | Withstand 1000 volt spike per EN 61000 | |
Response Time | < 380 mSec |
Housing
|
PVC
|
|
Sensor Diaphragm
|
Ceramic
|
|
O-ring Seals
|
FKM (Standard) or EPDM (Optional)
|
Pressure Connection
|
Open Face
|
|
Ingress Protection
|
IP 68
|
|
Overpressure Limit
|
6x Minimum (≥ 134 Ft. range = 3.5x)
|
|
Weight | 1.4 lbs. (without cable) |
EA
|
Special Calibration Run
|
|
PW
|
Calibrate Transmitter with Meter / Display
|
|
DV | SharkCage Removable Protective Cage |
DL-AC-S4NE
The Viatran DL meter boasts specifications and functionality that make it one of the most advanced process meters available.